1720E Low Range Process Turbidimeter and sc100 Controller with LonWorks/Modbus

GENERAL

The turbidity monitoring system shall include at least one Turbidimeter and one interface unit. The system shall be capable of functioning as a single sensor system and also be easily expanded up to two turbidimeters per interface unit. The connections between the turbidimeter and interface unit will include plug & play connections.

TURBIDIMETER

The turbidimeter shall measure turbidity in the range of 0.001-100 NTU and be a

microprocessor-based, continuous-reading, on-line nephelometric instrument

meeting all design and performance criteria specified by USEPA method 180.1.

Light shall be directed through the surface of the sample and the detector shall be immersed in the sample, eliminating glass windows and flow cells. Optical components shall be mounted in a sealed head assembly that can be removed easily for calibration/service, without disturbing sample flow. The turbidimeter body shall be constructed of corrosion-resistant polystyrene, and shall include an internal bubble removal system to vent entrained air from the sample stream.

The turbidimeter shall offer the choice of formazin-based (20 or 1 NTU) or instrument comparison-based calibration methods. Accuracy shall be ±2% of reading or ±0.015 NTU (whichever is greater) from 0 to 40 NTU; ±5% of reading from 40 to 100 NTU. Displayed resolution shall be 0.0001 NTU from 0 to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU and repeatability shall be better than ±1.0% of reading or 0.002 NTU (whichever is greater). User selectable signal averaging, bubble removal, alarm and recorder output hold, and self-test diagnostics shall be provided. All turbidimeters on the network shall have the option for MODBUS/RS232, MODBUS/RS485, LonWorks serial input/output capability for two-way communication to a computer or have wireless downloading capability through the IR Port located on the interface unit to download and print real-time turbidity data, calibration history, and current set points in a CSV format.

INTERFACE MODULE

The Interface unit shall allow operators to control sensor and interface functions

with user-friendly, menu-driven software, and shall provide data logging of

measurement data from up to two turbidimeters for 15 minutes, 1 hour, 24 hours, 30 days, or 180 days and the capability to transfer data to a computer or printer via direct MODBUS communications or directly into a Personal Digital Assistant (PDA) via a wireless IR Port. The interface unit will also have a built-in data logger with the storage capacity to store data on 15-minute intervals for up to 6 months with two sensors per controller. Each interface will also include two analog outputs and 3 un-powered SPDT alarm contacts. The interface unit and the DC power supply shall be housed in a NEMA-4X (indoor) industrial metal/plastic enclosure, and the power supply shall automatically accept input in the range of 100 to 230 Vac, 50/60 Hz.

SAFETY AND ELECTRICAL DESIGN STANDARDS

All system components are ETL listed to UL 61010A-1, certified by ETL to CSA C22.2 No. 1010.1, and CE certified by manufacturer for safety to EN 61010-1. For EMC immunity and emissions, system components are CE certified by manufacturer to EN 61326 (industrial levels), for North America to FCC Part 15, and Canadian Interference-Causing Equipment Regulation ICES-003, and for rest of world to CISPR 11 Class A levels.

 

 

责任编辑: admin
[ 产品特性 ] [ 技术指标 ] [ 选件&附件 ] [ 获奖&评测 ] [ 订购指南 ]

相关产品
2100N/AN系列实验室浊度仪 - 04-12-13 - 阅: 85313
表面散射高量程浊度仪 - 04-12-12 - 阅: 54101
1720E浊度测量系统 - 04-12-11 - 阅: 148082
GLI Accu4低量程浊度仪 - 04-04-29 - 阅: 55215
1720D 在线式浊度仪 - 02-10-19 - 阅: 89892
新闻
哈希1900C经济型便携式浊度仪问世 - 08-08-28 - 阅: 2205
热销产品 1720e Amtax 2100N 试剂 dr2800 dr5000 CL17 DREL2800 sc100 - 07-01-09 - 阅: 8158
软件升级 1720 E / FT660 sc 的量程浊度仪 - 07-01-04 - 阅: 32592
Txpro-2 WP系列浊度仪停产 - 05-08-05 - 阅: 43210
实验室浊度计的读数高于在线监测浊度仪,哪一个仪器是正确的 - 05-02-23 - 阅: 35589
1720C 浊度分析仪显示的结果为负数 - 05-02-23 - 阅: 29275
1720C/SS6浊度分析仪和旧式CL-17输出的结果不在4-20mA的正常范围内 - 05-02-23 - 阅: 33004
1720C SS6浊度分析仪\旧式CL-17键盘被锁定并且不能重启 - 05-02-23 - 阅: 34003
2100N/2100AN台式浊度仪中没有样品时读数不为0.000 - 05-02-23 - 阅: 31302
2100N/2100AN浊度仪自动重启 - 05-02-23 - 阅: 36785
2100N/2100AN台式浊度仪上显示E09错误信息 - 05-02-23 - 阅: 30863
2100P浊度仪显示闪烁的9.99或99.9 - 05-02-23 - 阅: 29994
2100P型浊度仪的出厂校准值 如何恢复? - 05-02-23 - 阅: 28938
使用Cal-Cube对Accu4浊度仪进行标定(校准) - 05-02-23 - 阅: 24240
Accu4或Steady Stream 4 浊度仪需同时更换LED光源 - 05-02-23 - 阅: 25053
Accu4浊度仪和Steady Stream 4 稳定流量浊度仪 - 05-02-23 - 阅: 25326
1720D浊度仪无法将传感器安装到Aquatrend网络 - 05-02-23 - 阅: 27713
HACH袖珍浊度仪停产通知 - 05-01-18 - 阅: 26662

Email给朋友 打印本文
版权所有 安恒公司 Copyright © 2000 AnHeng.com All Rights Reserved  
北京市海淀区车公庄西路乙19号 华通大厦B座北楼12层 (邮编 100048)
电话:北京(010-88018877) 上海(021-63176770)